User Entered values with the Pressure Module + Pump as Source and user entry as Measure.


Scenario:

User needs to run a calibration in which they are sourcing Pressure via a pump and pressure module but capturing a manually typed read out for calibration As Found/As Left.


Example:

The main Calibration Points tab should hold the Signal Types and ranges for display on a report like a calibration certificate.

E.g. 0 to 100 kPa for both input and output, One or more test points, as required.


The "DPC Settings" or "Group Options" area should use Pressure as Source mode.

For measure mode there are some options such as

  • RTD
    • requires a Probe Type and # of Wires
    • requires Unit Type of "Custom" and Custom Base Unit of DEGC, Custom Base Range of Low 0.000, High 100.000, or similar
  • TC
    • requires a Probe Type and CJC Default
    • requires Unit Type of "Custom" and Custom Base Unit of DEGC, Custom Base Range of Low 0.000, High 100.000, or similar
  • DC Voltage
    • requires Unit Type of "Custom" and Custom Base Unit of MV, Custom Base Range of Low 0.000, High 100.000, or similar

The idea here is that we can allow the reading to move forward even when the calibrator is ignoring the actual measured value, and enforcing that the User enter the reading value.


in all cases the output/measure setting of Reading Method is critical to allowing the 754 to capture the manually entered output. It must be set to UEV for User Enter Values.


Example using DC Voltage example